The device was returned for investigation.Upon evaluation of the device, foreign matter was adhered to the forceps.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
|
|
This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation.A review of the device history record found no deviations, that could have caused or contributed to the foreign matter that had adhered to the forceps.The investigation determined, that the cups failed to open and close, because foreign materials had adhered to the linkage mechanism.A definitive root cause for this issue was not established.However, it is probable that issue occurred, because of insufficient cleaning.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu), which state the following: ¿reprocess the instrument immediately, after use by immersing it in a neutral, low-foaming, medical-grade detergent solution.Then following the cleaning and sterilization procedures in this chapter.Failure to reprocess the instrument immediately after use, or using another type of detergent may cause corrosion at the instrument¿s distal end.This could cause components to fall off, during use and may interfere with operation of the cups.When inserting the instrument into the endoscope, hold it close to the biopsy valve or t-plug and keep it as straight as possible relative to the biopsy valve or t-plug.Otherwise, the insertion portion could be damaged.Lubrication: 1.Immerse the insertion portion in the lubricant for 2 to 3 seconds.2.Remove the instrument from the lubricant.3.Operate the slider to open and close the cups two or three times.4.Wipe the exterior of the instrument with a clean, dry lint-free cloth and allow the instrument to air dry¿.Olympus will continue to monitor field performance for this device.
|