• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. BIOPSY FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. BIOPSY FORCEPS Back to Search Results
Model Number FB-19K-1
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
A user facility reported that the biopsy forceps cup failed to open.The event occurred during preparation for use and a similar device was used to complete the operation.During a standard service inspection of the customer returned device, foreign matter was adhered to the forceps.This report is to capture the reportable malfunction found at inspection.There was no patient harm or user injury reported due to the event.
 
Additional Manufacturer Narrative
The device was returned for investigation.Upon evaluation of the device, foreign matter was adhered to the forceps.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation.A review of the device history record found no deviations, that could have caused or contributed to the foreign matter that had adhered to the forceps.The investigation determined, that the cups failed to open and close, because foreign materials had adhered to the linkage mechanism.A definitive root cause for this issue was not established.However, it is probable that issue occurred, because of insufficient cleaning.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu), which state the following: ¿reprocess the instrument immediately, after use by immersing it in a neutral, low-foaming, medical-grade detergent solution.Then following the cleaning and sterilization procedures in this chapter.Failure to reprocess the instrument immediately after use, or using another type of detergent may cause corrosion at the instrument¿s distal end.This could cause components to fall off, during use and may interfere with operation of the cups.When inserting the instrument into the endoscope, hold it close to the biopsy valve or t-plug and keep it as straight as possible relative to the biopsy valve or t-plug.Otherwise, the insertion portion could be damaged.Lubrication: 1.Immerse the insertion portion in the lubricant for 2 to 3 seconds.2.Remove the instrument from the lubricant.3.Operate the slider to open and close the cups two or three times.4.Wipe the exterior of the instrument with a clean, dry lint-free cloth and allow the instrument to air dry¿.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOPSY FORCEPS
Common Device Name
BIOPSY FORCEPS
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14427387
Report Number8010047-2022-08368
Device Sequence Number20205706
Product Code EOQ
UDI-Device Identifier04953170029608
UDI-Public04953170029608
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K962555
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFB-19K-1
Device Lot NumberK1819-1400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer08/03/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/19/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-