The device was returned for investigation.Upon evaluation of the device, the reported issue was confirmed.In addition, the wire coating was torn and the broken part was burnt and melted.When the outer diameter of the knife wire was measured, there was no abnormality.However, when the length of the knife wire and the wire coating was measured, the wire coating was missing 5.0 to 7.0 mm from the tip side.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The cutting wire was broken.Investigation was carried out to confirm the broken portion.The coated portion of the cutting wire was torn, and the broken portion was scorched and melted.The outer diameter of the cutting wire was measured.The result indicated no abnormalities.The length of the cutting wire and the coated portion was measured.The distal side of the coated portion was missing approximately 5.0-7.0 mm.Other abnormalities that could lead to the breakage of the cutting wire were not confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications upon release.Based on the results of confirmation of the device and the investigation results in the past, a likely mechanism causing the broken cutting wire might be the following: 1.The cutting wire at a torn area of the coated portion came into contact with the distal end of the endoscope while the forceps elevator was raised.2.The output was activated in state of ¿1¿ description, and the cutting wire became hot instantly.This caused the cutting wire to break.The event can be prevented by following the instructions for use which state: ¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.¿ when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.¿do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.¿if you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.Olympus will continue to monitor field performance for this device.
|