Model Number 2353-4107
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that during a tibial nail procedure, the targeter would not allow the screw to engage the nail.The targeting assembly was removed and the screw was implanted free-hand.Surgery was completed successfully with a delay of approximately 10-15 minutes.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the returned device is conforming to specifications and fully functional.Functional check prior to use is required in the labelling ¿ such as ifu, op-tech and re-processing brochure.Simulation of a pre-operative functional check of returned part[s] with sample items revealed no deficiency in targeting accuracy.Reasons for mal alignment are various, e.G.[but not limited to] insufficient tightening of the nail holding screw, applying too much [axial] force on the construct during drilling or unsuitable bone geometry.However, in this case freehand locking was finally performed, which is always an option for distal locking as per operative technique, and procedure was completed successfully.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.With available information a real root cause could not be determined but it could not be excluded that it is rather related to a sub-optimal intraoperative procedure.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
 
Event or Problem Description
It was reported that during a tibial nail procedure, the targeter would not allow the screw to engage the nail.The targeting assembly was removed and the screw was implanted free-hand.Surgery was completed successfully with a delay of approximately 10-15 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGETING ARM TIBIA T2 ALPHA? TIBIA
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14427749
Report Number0009610622-2022-00187
Device Sequence Number5629143
Product Code HSB
UDI-Device Identifier07613327357080
UDI-Public07613327357080
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K191271
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2353-4107
Device Catalogue Number23534107
Device Lot NumberKME913322
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/23/2022
Supplement Date Received by Manufacturer07/26/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age29 YR
Patient SexFemale
-
-