| Model Number |
2353-4107 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that during a tibial nail procedure, the targeter would not allow the screw to engage the nail.The targeting assembly was removed and the screw was implanted free-hand.Surgery was completed successfully with a delay of approximately 10-15 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed, since the returned device is conforming to specifications and fully functional.Functional check prior to use is required in the labelling ¿ such as ifu, op-tech and re-processing brochure.Simulation of a pre-operative functional check of returned part[s] with sample items revealed no deficiency in targeting accuracy.Reasons for mal alignment are various, e.G.[but not limited to] insufficient tightening of the nail holding screw, applying too much [axial] force on the construct during drilling or unsuitable bone geometry.However, in this case freehand locking was finally performed, which is always an option for distal locking as per operative technique, and procedure was completed successfully.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.With available information a real root cause could not be determined but it could not be excluded that it is rather related to a sub-optimal intraoperative procedure.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
|
| |
|
Event or Problem Description
|
|
It was reported that during a tibial nail procedure, the targeter would not allow the screw to engage the nail.The targeting assembly was removed and the screw was implanted free-hand.Surgery was completed successfully with a delay of approximately 10-15 minutes.
|
| |
|