| Model Number |
4406 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Skin Inflammation/ Irritation (4545)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A photo was provided for the evaluation.Visual examination of the photo shows abrasions on the patient's skin which verifies the reported issue.Without the actual sample, a definitive root cause could not be confirmed as a defect of the clipper.Panasonic states the most probable root cause was caused by using the handle/blade with strong pressure to the skin, or by using the broken blade.A production record review for the batch/lot could not be completed as no information was available.An in-service was training was completed by the bd sales associate with the reporter's site.This failure will continue to be tracked and trended.
|
| |
|
Event or Problem Description
|
|
It was reported that there has been an increase in clipper razor burns.Verbatim: indicates that they have seen an increase in clipper razor burns and asked for a refresher on proper and safe use for the staff who clips patients prior to surgery.
|
| |
|