• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENUSA DE MEXICO S.A. DE C.V. TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 65MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENUSA DE MEXICO S.A. DE C.V. TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 65MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number PN-004 065
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One 65 mm tacticath quartz contact force ablation catheter was received for evaluation.Electrodes 1-4 met specifications during electrical testing with no open or short circuits detected.The device did not meet specifications during a shaft leak test.Further investigation revealed the pi irrigation tubing had been fractured near the irrigation tubing transition near the deflectable portion of the catheter shaft, consistent with the failed shaft leak test, fluid ingress, and with the reported event.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the fractured irrigation tubing could not be conclusively determined.
 
Event or Problem Description
This report is to advise of an event noted during analysis which revealed a leak in the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 65MM
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX  32470
Manufacturer (Section G)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX   32470
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14428314
Report Number9680001-2022-00020
Device Sequence Number5628639
Product Code OAE
UDI-Device Identifier07640157990033
UDI-Public07640157990033
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P130026
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/25/2022
Device Model NumberPN-004 065
Device Catalogue NumberPN-004065
Device Lot Number7525979
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Initial Report FDA Received Date05/18/2022
Date Device Manufactured10/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-