| Model Number |
7278 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
|
| |
|
Event or Problem Description
|
|
The customer performed a blood glucose testing with the contour next link 2.4 meter and obtained a reading of 43 mg/dl at 1:00 p.M.The meter automatically marked it as a control test, and so the reading will be displayed as a control result when accessing the meter's memory.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer did not return the suspected device for evaluation.Therefore, the in-house contour next link 2.4 meter was tested with the in-house contour next test strips from lot # 1mpeh41a using blood sample, which gave a satisfactory performance.The blood glucose readings obtained during in-house testing were properly marked as blood tests in the meter's memory.Additionally, a device history record was reviewed for the suspected contour next link 2.4 meter and no manufacturing anomalies were found.
|
| |
|