U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE MOBILE POWER UNIT, NA; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE MOBILE POWER UNIT, NA; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 107754
Medical Device Problem Code Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Event or Problem Description
It was reported that the patient underwent a system controller exchange due to damage to bend relief.Further review of the log file revealed no external power alarm that activated on (b)(6) 2022 at 00:07 due to ac power disconnect while the device was connected to a mobile power unit.
 
Additional Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of a no external power alarm was confirmed via the log file review.The log file spanned approximately 2 days ((b)(6) 2022 per the timestamp).A no external power alarm activated on (b)(6) 2022 at 00:07 due to the rsoc and bus voltage simultaneously diminishing to ~0v with a loss of ac power while connected to the mpu.The backup battery was able to provide power to the pump during the alarm.The alarm resolved shortly after reconnecting to a power source.No other notable alarm was observed.The mobile power unit, serial (b)(6), was not returned for analysis.Additional information on 02jun2022 stated that the mpu have a v-lock power cord.The cause of the no external power alarm cannot be determined.A root cause of the reported event was not determined through this analysis.Device history records were reviewed and showed no deviations from manufacturing or qa specifications.The patient handbook cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.Heartmate iii instructions for use section 7-¿alarms and troubleshooting¿ and heartmate iii patient handbook section 5-¿alarms and troubleshooting¿ explain how to properly interpret and troubleshoot alarms including no external power.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE MOBILE POWER UNIT, NA
Common Device Name
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key14428425
Report Number2916596-2022-10686
Device Sequence Number6634796
Product Code DSQ
UDI-Device Identifier00813024013006
UDI-Public00813024013006
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P160054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number107754
Device Catalogue Number107754
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/20/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age51 YR
Patient SexFemale
Patient Weight112 KG
-
-