Model Number TFNT40
Medical Device Problem Codes Application Program Problem: Power Calculation Error (1473); Malposition of Device (2616)
Health Effect - Clinical Codes Visual Impairment (2138); Visual Disturbances (2140)
Date of Event 10/21/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient experienced double vision and problems focusing which was discovered the same day as it was implanted.The surgeon's opinion the axis was off for the iol.The surgeon was rotated the iol.Also the patient had prk (photorefractive keratectomy) procedure.The iol was exchanged for another lens model 5 months following the initial implant procedure.There was no patient harm reported and symptoms were resolved.
 
Additional Manufacturer Narrative
The customer indicated the use of qualified associated products.Information was provided that the company model (2.25cyl), 17.0 diopter (add power +2.2/+3.2) lens was replaced with a company (2.25cyl), 17.5 diopter, (1.5 extended depth of focus) lens with a clinical reason for the exchange of "diplopia." each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Additional information was provided by a facility representative that in the surgeon's opinion, the axis was off for the iol.Facility representative also reported there was no patient harm and the patient's issue have resolved since the lens exchange.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The lens was returned wrapped in surgical gauzes.Solution was dried on the lens.The lens was cut into multiple pieces, typical of insertion and removal.Power and resolution testing could not be conducted due to the extensive optic damage.The product investigation could not identify a root cause for the reported complaint.Power and resolution testing could not be conducted due to the extensive optic damage.The power and resolution of each lens is 100% evaluated during the manufacturing process to determine acceptability per model and diopter.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14428614
Report Number1119421-2022-01057
Device Sequence Number5628648
Product Code MFK
UDI-Device Identifier00380652415996
UDI-Public00380652415996
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberTFNT40
Device Catalogue NumberTFNT40U170
Device Lot Number12748853
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer06/29/2022
11/16/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date07/25/2022
12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D.; MONARCH III IOL DELIVERY SYSTEM, INJECTOR.; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-