| Model Number |
AVEA STANDARD |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Vyaire medical file identification: (b)(4).The device was not returned for further evaluation; therefore a definitive root cause could not be determined and corrective or preventive action is not indicated at this time.However, the customer was provided with calibration codes.
|
| |
|
Event or Problem Description
|
|
The customer reported to vyaire medical that the avea ventilator was experiencing a positive end-expiratory pressure (peep) instability and requested a peep calibration code.The event happened during set up prior patient use.Furthermore, there was no patient associated with the reported event.
|
| |
|