MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCUSTAT; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCUSTAT; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-6110
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during photoselective vaporization of the prostate this fiber exhibited an internal break approximately two to three inches above the handpiece.A new fiber was used to complete the procedure.There were no patient complications.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this fiber was thoroughly analyzed.Visual inspection noted the fiber body was kinked and exhibited a break.The fiber was functionally tested with the helium-neon (hene) laser fixture.The testing identified the majority of the output escaped from the fracture location.The fiber body break was determined to be the likely result of excessive manipulation/inappropriate technique during fiber setup.As a result, the probable cause of unintended use error caused or contributed to event was assigned to this investigation.
 
Event or Problem Description
It was reported that during photoselective vaporization of the prostate this fiber exhibited an internal break approximately two to three inches above the handpiece.A new fiber was used to complete the procedure.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUSTAT
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key14429535
Report Number2124215-2022-17099
Device Sequence Number15091083
Product Code GEX
UDI-Device Identifier00878953000121
UDI-Public00878953000121
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K924644
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2022
Device Model Number0010-6110
Device Catalogue Number0010-6110
Device Lot Number0023984809
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/09/2022
Supplement Date Received by Manufacturer06/10/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1