Model Number 71453141
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after a tka on (b)(6) 2019, the femur had a streak down at the medial condyle and the lgn xlpe cr insert sz 1-2 9 mm was broken.The patient required a revision surgery and the current status of the patient is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation but the pictures were reviewed, and the fracture was confirmed.The clinical/medical investigation concluded that, based on the single photo image provided, the root cause of the reported event was most likely an insert disassociation with subsequent femoral condyle and tibial baseplate articulation; however, this cannot be definitively concluded.It is unknown if the patient experienced any trauma prior to the reported event.The patient impact beyond the suspected insert disassociation with metal-on-metal articulation with subsequent revision could not be determined.No further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed revealed that the implant can break or become damaged as a result of strenuous activity or trauma and has a finite expected service life and may need to be replaced in the future.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury or surgical technique.The contribution of the device to the reported incident could be corroborated as the device was broken and it was required a revision surgery to treat this event.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN XLPE CR INSERT SZ 1-2 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14429775
Report Number1020279-2022-02391
Device Sequence Number5604893
Product Code JWH
UDI-Device Identifier00885556255407
UDI-Public00885556255407
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71453141
Device Catalogue Number71453141
Device Lot Number18HM21517
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer06/10/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date06/14/2022
Date Device Manufactured08/29/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GII CM TIB SIZE 2 {} RIGHT, LN: 19JM06744; GII OVAL RESURFACING PAT 29MM, LN: 19HM02937
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-