ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA22-80/I16-40
Medical Device Problem Codes Material Puncture/Hole (1504); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Aneurysm (1708); Failure of Implant (1924)
Date of Event 04/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.Devices remain implanted.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft.Approximately three (3) years post initial procedure, the patient presented with aneurysm sac enlargement (unknown diameter) with an endoleak (at indeterminate origin).The physician elected to treat the patient by implanting one (1) endurant (non-endologix) device below the inferior mesenteric artery, and two (2) excluder (non-endologix) devices in the right and left common iliac arteries on (b)(6) 2022.The endoleak was successfully resolved.The final patient status was not reported.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft.Approximately three (3) years post initial procedure, the patient presented with aneurysm sac enlargement (unknown diameter) with an endoleak (at indeterminate origin).The physician elected to treat the patient by implanting one (1) endurant (non-endologix) device below the inferior mesenteric artery, and two (2) excluder (non-endologix) devices in the right and left common iliac arteries on (b)(6) 2022.The endoleak was successfully resolved.The final patient status was not reported.Additional information: the indeterminate endoleak was determined to be a type 3b endoleak.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the indeterminate endoleak was determined to be a type 3b endoleak of the bifurcated stent graft.The sac growth of 9.5mm was also confirmed.This is consistent with the reported adverse event/incident.Contributing factors for this event are as follows, which were off-label conditions: the aortic neck size was 17mm (when it should have been 18-22mm), and intended aortic vessel diameter for a 22mm bifurcated stent graft should be 18-20mm (but the distal aortic diameters on the pre-planning image were 14.4mm, 13mm and 9x18.7mm).The procedure-related harms, device, user, procedure, or anatomy-relatedness of this complaint could not be determined with the medical records available for review.The final patient status remains unknown.The final patient status was not made available to endologix.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections: b5: describe event or problem - updated.G3 awareness date.H6 investigation finding codes; remove 3233.H6 investigation conclusion codes; remove 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key14429855
Report Number2031527-2022-00100
Device Sequence Number6530481
Product Code MIH
UDI-Device Identifier00818009015006
UDI-Public(01)00818009015006(17)190625
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 05/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/25/2019
Device Model NumberBEA22-80/I16-40
Device Lot Number1951893-002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/08/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/18/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0007-2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-