| Catalog Number |
2N1191 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a non-dehp y-type catheter extension set leaked.The event was further described as; during infusion, "cytotoxic medication leakage onto a patients clothing".The cytotoxic medication was docetaxel.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.All components were correctly placed and according to specifications.A functional testing was performed including clear passage and pressure testing; and the results were satisfactory.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|