MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 68 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 68 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 68LP-250
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 12/06/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported left side capsular contracture, baker grade iv.Device has been explanted.Healthcare professional reported treatment with "bra band/massage.".
 
Additional Manufacturer Narrative
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture, baker grade iv" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iv.
 
Additional Manufacturer Narrative
Device evaluation:visual analysis of the returned device identified: opening and white particle inside device.Leak test was performed and found opening in the anterior.A microscopic analysis was performed which identify opening striated in the anterior.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a striated opening in the anterior side assessed as surgical damage.Opening.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture, baker grade iv.Device has been explanted.Healthcare professional reported treatment with "bra band/massage.".
 
Event or Problem Description
Healthcare professional reported left side capsular contracture, baker grade iv.Device has been explanted.Healthcare professional reported treatment with "bra band/massage.".
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b3.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture, baker grade iv.Device has been explanted.Healthcare professional reported treatment with "bra band/massage.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 68 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key14430902
Report Number9617229-2022-08014
Device Sequence Number6428769
Product Code FWM
UDI-Device Identifier10888628000568
UDI-Public10888628000568
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2025
Device Model Number68LP-250
Device Catalogue Number68LP-250
Device Lot Number3515795
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer06/02/2022
12/19/2022
01/09/2023
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date06/28/2022
01/04/2023
01/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexFemale