Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.The icd interrogation revealed the eos battery status, detected after explantation of the device on (b)(6), 2021.The device was implanted for approximately 70 months.An amount of 29 charging cycles was documented.The memory content of the icd was inspected, revealing a normal current consumption of this device.However, the amount of charge taken from the battery was verified and the battery condition was not as expected.Therefore, the icd was opened and the inner assembly was inspected.The visual inspection of the inner assembly showed no anomalies.In a next step, the electronic module was attached to an external power supply and the eos status was removed with a technical programmer to check the functionality of the electronic module.The measured current consumption was normal and as expected.There was no indication of a malfunction, particularly all therapy functions were available and worked as expected.Therefore, the battery was disconnected from the electronic module and sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the battery manufacturing.No anomalies were documented during the production process, associated with this battery.The visual inspection of the battery did not reveal any external signs of damage.The voltage measurement confirmed the battery depletion and the microcalorimetry analysis showed an elevated heat dissipation.At a next step, the battery was opened for destructive analysis and the inner assembly was inspected.The inspection of the inner assembly identified a short circuit within the battery, which led to the elevated internal self-depletion and, as a result, to the clinical observation.In conclusion, the device was implanted for about 70 months.The therapeutic functionality of the device was tested with an attached external power supply and proved to be fully functional.Further investigations revealed an increased internal self-depletion within the battery that contributed to the clinical observation.
|