| Model Number |
M00573020 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Approximated based on the date the manufacturer became aware of the event.(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a jagtome rx 49 was used in a procedure.The exact procedure date was unknown.During the procedure, the cutting wire of the jagtome rx 49 broke at the time of cutting.There were no reported patient complications as a result of this event.No further information has been obtained despite good faith efforts.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a jagtome rx 49 was used in a procedure.The exact procedure date was unknown.During the procedure, the cutting wire of the jagtome rx 49 broke at the time of cutting.There were no reported patient complications as a result of this event.No further information has been obtained despite good faith efforts.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: medical device problem code a0401 captures the reportable event of cutting wire broken.Block h10: the returned jagtome rx 49 was analyzed, and a visual evaluation noted that the cutting wire was broken and kinked at the proximal pierced hole, consistent with the findings when the device was observed under magnification.The ends of the cutting wire were also blackened.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken and blackened.The cutting wire being blackened indicates that the device was energized.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.Additionally kinking/bending the cutting wire can lead to break it.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
| |
|