• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. SMARTLOAD; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. SMARTLOAD; INTRAOCULAR LENS Back to Search Results
Model Number GCB00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the intraocular lens (iol) had a defect.Therefore, the surgeon cut it after implantation and implanted a new iol.Through follow-up, it was learned that there was a round approximately 1mm sized flaw on the optic that was seen during implantation.There was a 15 minute delay in the procedure, where the incision was enlarged, and the iol was removed and replaced with a new lens.No other medical interventions were necessary.The patient's pre operative visual acuity was 0.1 and the patient's post operative visual acuity was 0.1.No further information was provided.
 
Additional Manufacturer Narrative
Information unknown/ not provided.Per eu regulation 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.(b)(6).Initial reporter - first/given name: unknown/not provided.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Section d9 - device available for evaluation? yes.Section d9 - date returned to manufacturer: 26th may 2022.Section h3 - device evaluated by manufacturer? yes.Device evaluation: the complaint lens was received inside of a specimen vial.No additional materials were received as part of this return.Visual inspection under magnification revealed that the lens was received cut in half (one half missing).The lens was cleaned, and no further issues could be identified with the lens.Based on the return condition of the lens, no further product evaluation could be performed.As a result of the investigation, there is no indication of a product quality deficiency.The complaint issue (dc-lens damaged) was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTLOAD
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14431168
Report Number3012236936-2022-01259
Device Sequence Number20440162
Product Code HQL
UDI-Device Identifier05050474620636
UDI-Public(01)05050474620636(17)250201
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2025
Device Model NumberGCB00
Device Catalogue NumberGCB0000215
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer07/01/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date07/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-