| Model Number |
GCB00 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the intraocular lens (iol) had a defect.Therefore, the surgeon cut it after implantation and implanted a new iol.Through follow-up, it was learned that there was a round approximately 1mm sized flaw on the optic that was seen during implantation.There was a 15 minute delay in the procedure, where the incision was enlarged, and the iol was removed and replaced with a new lens.No other medical interventions were necessary.The patient's pre operative visual acuity was 0.1 and the patient's post operative visual acuity was 0.1.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Information unknown/ not provided.Per eu regulation 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.(b)(6).Initial reporter - first/given name: unknown/not provided.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9 - device available for evaluation? yes.Section d9 - date returned to manufacturer: 26th may 2022.Section h3 - device evaluated by manufacturer? yes.Device evaluation: the complaint lens was received inside of a specimen vial.No additional materials were received as part of this return.Visual inspection under magnification revealed that the lens was received cut in half (one half missing).The lens was cleaned, and no further issues could be identified with the lens.Based on the return condition of the lens, no further product evaluation could be performed.As a result of the investigation, there is no indication of a product quality deficiency.The complaint issue (dc-lens damaged) was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|
|