| Model Number |
SSLP-360 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Biocompatibility (2886)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
08/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported capsular contracture, baker grade iii against right device.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified flat creases and one curved opening.A weight test of the device was verified and the device was within specification.A microscopic analysis was performed which identified one opening striated.Based on the device analysis the final assessment is: one opening striated in the side radius assessed as surgical damage.Additional, changed, and/or corrected data: b5, d6b, d9, h3, h6.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|