Model Number TMICL12.6
Medical Device Problem Codes Misfocusing (1401); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The reporter indicated the surgeon implanted a 12.6mm tmicl12.6 implantable collamer lens, -11.00/+3.0/136 (sphere/cylinder/axis), in the patients right eye (od) on (b)(6) 2022.The lens was explanted on (b)(6) 2022 due to lens rotation not associated to a low vault and patient experienced a refractive surprise.The lens was replaced with a longer lens and the problem was resolved.The cause of the event was unknown.
 
Additional Manufacturer Narrative
Work order search: no similar complaint was reported for units within the same lot.Claim # (b)(4).
 
Additional Manufacturer Narrative
H3: device evaluation: the lens was returned in liquid in a vial.Visual inspection found no visible damage to the lens.Dimensional inspection found the lens to be within specifications.Functional inspection found the returned lens did not meet original values measured at the time of manufacturing.Claim #(b)(4).
 
Additional Manufacturer Narrative
H6 - type of investigation 3331 - device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key14431984
Report Number2023826-2022-01576
Device Sequence Number17749507
Product Code QCB
UDI-Device Identifier00841542107020
UDI-Public00841542107020
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberTMICL12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer09/16/2022
09/16/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date09/19/2022
09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL SFC-45 - LOT #UNK; FOAM TIP PLUNGER MODEL FTP - LOT #UNK; INJECTOR MODEL MSI-PF - LOT #UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexMale
-
-