• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL LTV 1200 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL LTV 1200 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number LTV 1200
Medical Device Problem Code Pressure Problem (3012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The customer reported to vyaire medical problems with lap top ventilator 1200.The respiratory therapist noted a high pressure reading, but inaccurate.The vent is also due for pm.Furthermore, the customer confirmed that alternative ventilation was provided to the patient.Respiratory therapist was able to switch the ventilator and there was no patient harm reported associated with the event.
 
Additional Manufacturer Narrative
The suspect device was returned and evaluation is anticipated but not yet begun.Once a final investigation is complete, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Results of investigation: vyaire medical received the suspect device for evalution.Event trace download reveals several high-pressure alarm condition occurred but was unable to duplicate during bench testing at service.All testing was performed with a good known test ac adapter, patient circuit, and test lung.Vent passed 48 hours extended tests at customer settings with no unusual alarms or conditions.Vent passed ts final checkout test (service manual 33132-001 version b )which includes many alarm and ventilation functions.Process vent thru service and perform complete checkout to meet all factory specification and standards.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LTV 1200 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key14432075
Report Number2021710-2022-16049
Device Sequence Number15069262
Product Code CBK
UDI-Device Identifier10190752115637
UDI-Public(01)10190752115637(11)20180110
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K060647
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLTV 1200
Device Catalogue Number18888-201-99
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer05/05/2023
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date05/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-