• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CTS22, 12X100 KII SLEEVE ZTHR 12/BX; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CTS22, 12X100 KII SLEEVE ZTHR 12/BX; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number CTS22
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Procedure performed: oesophagus resection.Event description: [translation]: today, during the oesophagus resection, we saw that our port had been broken.We thought this was quite dangerous.Fortunately, no plastic had broken off, but the plastic had bent.Additional information received by email from applied medical representative on (b)(6) 2022: patient status: is ok.The trocar was inserted between the thoracal bone.There was no difficulty during insertion.The trocar was not used with a robot, laparoscopic.Instruments inside of the cannula at the time of the incident: [name] grasper, needle holder.Patient status: no patient injury has been reported.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of a fractured cannula tip.Based on the condition of the returned unit, it is possible that the cannula tip fractured due to instrumentation coming in contact with the tip during the procedure.It is also possible that the cannula was more susceptible to breakage.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Event or Problem Description
Procedure performed: oesophagus resection.Event description: [translation]: today, during the oesophagus resection, we saw that our port had been broken.We thought this was quite dangerous.Fortunately, no plastic had broken off, but the plastic had bent.Additional information received by email from applied medical representative on 10may22: patient status: is ok.The trocar was inserted between the thoracal bone; there was no difficulty during insertion.The trocar was not used with a robot, laparoscopic instruments inside of the cannula at the time of the incident: [name] grasper, needle holder.Patient status: no patient injury has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CTS22, 12X100 KII SLEEVE ZTHR 12/BX
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key14432077
Report Number2027111-2022-00602
Device Sequence Number20449459
Product Code GCJ
UDI-Device Identifier00607915123734
UDI-Public(01)00607915123734(17)250310(30)01(10)1444762
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K060096
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2025
Device Model NumberCTS22
Device Catalogue Number101472328
Device Lot Number1444762
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2022
Supplement Date Received by Manufacturer05/02/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date06/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LIGASURE MARYLAND, GRASPER, NEEDLE HOLDER.
Patient SexPrefer Not To Disclose
-
-