| Model Number |
115-547 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/12/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: underweight, broken shell, missing a piece of shell (25-50%), wear abrasion, and crease fold.A microscopic analysis was performed which identified: sharp broken with stress marks near of the broken site, this condition was called surgical impact.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: sharp broken on anterior assessed as surgical impact.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.The device has been explanted.
|
| |
|