| Model Number |
SC-8352-70 |
| Medical Device Problem Codes |
Unexpected Therapeutic Results (1631); Migration (4003)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
04/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient was experiencing inadequate stimulation of the targeted pain area; bilateral back and legs.Multiple reprogramming attempts were made, however, only the left inner thigh was able to obtain coverage.An x-ray was performed which confirmed that the lead had migrated.The patient underwent a lead revision procedure where the lead was explanted and replaced.The patient is expected to fully recover.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was experiencing inadequate stimulation of the targeted pain area; bilateral back and legs.Multiple reprogramming attempts were made, however, only the left inner thigh was able to obtain coverage.An x-ray was performed which confirmed that the lead had migrated.The patient underwent a lead revision procedure where the lead was explanted and replaced.The patient is expected to fully recover.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned lead was analyzed; passed all tests performed and exhibited normal device characteristics.Engineers have concluded that the reported event of lead migration, resulting in inadequate stimulation is a known inherent risk with implanting a pulse generator as part of a system to deliver spinal cord stimulation.
|
| |
|