| Model Number |
DFR00V |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Disturbances (2140)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Health care provider reported they explanted an intraocular lens (iol) due to patient experiencing intolerable dysphotopsia.The iol was replaced with a non-johnson & johnson lens of the same diopter.The problem was noticed during a post-op exam.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Pt info: unknown, information was requested but not provided.Date of event: the exact date is unknown, the best estimate is between the implant date (b)(6) 2022 and the explant date (b)(6) 2022.The device has not returned for evaluation.Therefore, a visual analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain the missing information.However, to date it has not been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: jun 7, 2022.Section h3: evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the complaint lens was received cut into three pieces.The lens was cleaned and, no issues that could contribute to a poor visual outcome could be identified.Based on the return condition of the lens, no further product evaluation could be performed on the lens.A product quality deficiency could not be determined.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
|
| |
|
Search Alerts/Recalls
|