| Medical Device Problem Codes |
Overheating of Device (1437); Device Remains Activated (1525)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Report confirmed.The device was tested and the maximum temperature was measured to be 170°f.The likely cause was identified as worn front and rear bearings.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with the report of overheating and runs unintentionally.No patient impact was reported.Repair request escalated to a product event based on reason for return.Additional information regarding the patient impact was being followed up from the facility.Date of notification: 28-mar-2022.
|
| |
|