| Model Number |
55890006545 |
| Medical Device Problem Code |
Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider via a manufacturer representative regarding a screw used for spinal surgery.Pre-op diagnosis was lumbar disc herniation.The procedure performed in the event was lumbar intervertebral disc herniation, posterior disc removal, and posterior lumbar screw fixation.It was reported that intra-operatively, screw plug slipped.The product came in contact with patient.There was no delay in the overall procedure time.There were no patient symptoms or complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis part # 55890006545; lot # h5712358 visual and optical inspection confirmed the threads on one side of the head of the screw have been damaged.The damage appears to be from the incorrect threading of the screw in the multi axial screw during use.H6: updated eval.Code result and eval.Code method post analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|