ALCON GRIESHABER AG BACKFLUSH HANDLE DSP; INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED, ACCESSORY Back to Search Results
Catalog Number 337.88
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Hematoma (1884); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A physician reported that an ophthalmic backflush came off in the patient's eye during retinal detachment surgery.The patient underwent to a medical intervention to remove the broken tip.Additional information has been requested none received till date.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information received from the physician indicating when picking up back flush.Tip came off while aspirating sub retinal fluid.It dropped into retinal tear hemorrhaged when picked up it with forceps.Hemorrhaged from the rupture on the lateral side of the eye, flowed into the sub retina and precipitated as a hematoma in the posterior pole of the eye.
 
Additional Manufacturer Narrative
Additional information is provided in section a.2, a.3, b.3., b.5., b.7., e.1.And h.6.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The investigation is completed.A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.Sample was not received by the investigation site for evaluation.The root cause cannot be identified conclusively because no picture or sample has been received.Investigation has been completed based on current information.Should sample returned, this file will be re-opened, and failure analysis will be performed.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BACKFLUSH HANDLE DSP
Common Device Name
INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED, ACCESSORY
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14433139
Report Number3003398873-2022-00033
Device Sequence Number17582143
Product Code HQE
UDI-Device Identifier07612717071414
UDI-Public07612717071414
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K884043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number337.88
Device Lot NumberF208945
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer06/06/2022
07/25/2022
Initial Report FDA Received Date05/18/2022
Supplement Report FDA Received Date07/04/2022
08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age58 YR
Patient SexMale
-
-