SYNTHES GMBH COLIBRI II HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.101 |
Device Problems
Noise, Audible (3273); Excessive Heating (4030)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/26/2022 |
Event Type
malfunction
|
Event Description
|
It was reported by the (b)(6) that during service and evaluation, it was determined that the small battery drive device had damaged mechanics, corroded bearing, insufficient/low power, excessive noise, produced heat and had cosmetic damage-worn coupling and damaged component.It was further determined that the device failed pretest for general condition and check power with power test bench.It was noted in the service order that the device was making lots of noise.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be traced to maintenance.Udi: (b)(4).
|
|