• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH COLIBRI II HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH COLIBRI II HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 532.101
Device Problems Noise, Audible (3273); Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/26/2022
Event Type  malfunction  
Event Description
It was reported by the (b)(6) that during service and evaluation, it was determined that the small battery drive device had damaged mechanics, corroded bearing, insufficient/low power, excessive noise, produced heat and had cosmetic damage-worn coupling and damaged component.It was further determined that the device failed pretest for general condition and check power with power test bench.It was noted in the service order that the device was making lots of noise.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be traced to maintenance.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLIBRI II HANDPIECE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key14433558
MDR Text Key292052815
Report Number8030965-2022-03286
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2022
Date Manufacturer Received05/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-