Model Number A6282
Medical Device Problem Codes Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that after an unspecified procedure it was noticed that a fragment of about 1-2mm of the ceramic insulation at the distal end of the hf hook electrode had broken off during the procedure and fallen into the patient.Since this was only noticed after the procedure, the fragment may have remained inside the patient.No further information was provided but there was no report about and adverse event or patient injury.
 
Additional Manufacturer Narrative
Device evaluation: the suspect medical device was returned to the manufacturer for evaluation/investigation.However, the exact cause of the reported event could not be conclusively determined.Based on its lot number, the electrode was manufactured in october 2019.The electrodes are designed for 400 cycles, which is approximately equivalent to two years of use.As a result, it is possible that the electrode was used longer than the service life it was tested for.It would also be conceivable that there was pre-existing damage due to improper handling during reprocessing.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.The case will be closed on olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-ELECTRODE, HOOK, 5 X 330 MM
Common Device Name
HF ELECTRODES, RIGID
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
BOWA-ELECTRONIC GMBH & CO KG
heinrich-hertz-strasse 4-10
gomaringen
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662955
MDR Report Key14434513
Report Number9610773-2022-00181
Device Sequence Number5029162
Product Code GCJ
UDI-Device Identifier04042761028868
UDI-Public04042761028868
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K923982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA6282
Device Catalogue NumberA6282
Device Lot Number942773
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer07/07/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-