| Model Number |
M0068403910 |
| Medical Device Problem Codes |
Break (1069); Failure to Fire (2610); Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on september 13, 2021 that a flexiva laser fiber was used in a flexible ureteroscopy with lithotripsy procedure for a stone treatment in the kidney performed on (b)(6) 2021.The laser console unit used was a dornier h20 laser console.During the procedure, the laser fiber was connected to the laser however, the green light did not appear at the tip.The laser fiber did not have any effective action on the kidney stones.The procedure was completed with another flexiva laser fiber.There were no patient complications reported as a result of this event.No further information has been obtained despite good faith efforts.Note: this event has been deemed reportable based on the investigation finding of laser fiber broken within the connector.
|
| |
|
Additional Manufacturer Narrative
|
|
The event was reported by a distributor, however the healthcare facility where the event occurred is as follows: (b)(6).(b)(4).Investigation results: the returned flexiva laser fiber was analyzed, and a visual evaluation noted that the laser fiber was returned in one piece.The laser fiber body measured 311.5 cm from the end of the connector to the end of the exposed glass tip.The exposed glass tip measured 3 mm and appeared in good condition with no signs of degradation.A functional evaluation revealed that there was no light present from the sma connector indicating a fiber connector fracture.No other problems with the device were noted.The reported event of the flexiva laser fiber aiming beam not visible was confirmed.The reported laser fiber aiming beam not visible is consistent with a fiber broken within the connector.It is likely that procedural handling of the device during set-up or use contributed to the fracture within the fiber connector.A broken fiber within the connector of the device can result in insufficient aiming beam and low laser energy output, up to a complete inability for the fiber to fire.Review and analysis of all available information indicated that the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
| |
|
Search Alerts/Recalls
|