TELEFLEX MEDICAL HUDSON ET TUBE, CUFFED, SPIRAL-FLEX 7.0; TUBE, TRACHEAL (W/WO CONNECTOR Back to Search Results
Model Number IPN044799
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.Current part number is not being manufactured currently, however, another part number from the same family was use for the "verification of failure mode reported in the current manufacturing process" and was conducted as follows: 80 samples were taken from the current production from the assembly,spiral-flex nasal tube,8.0 mm the samples were visually inspected and issue reported "leak - device leak - cuff" was not observed in the current manufacturing process.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported that: "following intubation of patient, it was noted that there was a leaky pilot cuff.Extra air was required several times during procedure, however procedure continued uneventfully and there was no issue for patient.' patient condition reported as "fine".
 
Additional Manufacturer Narrative
Qn#(b)(4).The customer returned one unit 5-12614 et tube, cuffed, spiral-flex 7.0 for investigation.The et tube was visually examined with and without magnification.Visual examination of the returned device that the pilot balloon was detached from the inflation line.The ifu for this product states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is a detected in any part of the system, the tube should not be used." "care should be taken to avoid damaging the thin-walled cuffs during intubation.If cuff is damaged, the tube should not be used." the ifu also states, "cuff pressure should be monitored routinely to assure that an adequate tracheal seal is maintained and that the cuff has not become over-inflated.The tracheal tube cuff should be slowly inflated with the minimum amount of air required to provide an effective tracheal seal.Do not inflate with a measured volume of air or by "feel" of pressure in the syringe since little resistance should be felt during inflation." the reported complaint was confirmed based upon the sample received.The pilot balloon was detached from the inflation line which would prevent the cuff from being able to inflate.The customer reported that there was a leak from the pilot balloon.It is possible that the pilot balloon leaked during use and detached during shipping since there was no mention that the pilot balloon detached during use.The manufacturing site was consulted for this complaint issue.It was confirmed that the pilot balloon was properly adhered to the inflation line.It appears that the pilot balloon was pulled too hard which caused it to detach from the inflation line.The tip of the inflation line that the pilot balloon attaches to was slightly damaged which supports this root cause.A device history record review was performed with no evidence to suggest a manufacturing related cause.All et tubes are 100% inspected for inflation and deflation at the time of manufacturing.It is unlikely that this type of damage was present at the time of manufacturing.Based on the condition of the sample received, it appears that unintentional user error caused or contributed to this event.
 
Event or Problem Description
It was reported that: "following intubation of patient, it was noted that there was a leaky pilot cuff.Extra air was required several times during procedure, however procedure continued uneventfully and there was no issue for patient.' patient condition reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON ET TUBE, CUFFED, SPIRAL-FLEX 7.0
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key14435258
Report Number3003898360-2022-00172
Device Sequence Number17554164
Product Code BTR
UDI-Device Identifier04026704397402
UDI-Public04026704397402
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K844296
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date10/21/2025
Device Model NumberIPN044799
Device Catalogue Number5-12614
Device Lot Number73K2000503
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date09/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-