Model Number 74122154
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Osteolysis (2377); Metal Related Pathology (4530)
Date of Event 12/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference number: case-(b)(4).
 
Event or Problem Description
It was reported that, after primary left hip bhr resurfacing surgery was performed on (b)(6) 2009, the patient experienced pain and subluxation of the femoral component, for which revision surgery was conducted on (b)(6) 2012 (covered under case-(b)(4)).Subsequently, the patient experienced cobalt and chromium toxicity, as well as acetabular osteolysis.This adverse event was treated with revision surgery on (b)(6) 2020.From all the components, only the stem was left in place.The system was replaced to a competitor´s tha.The patient tolerated the procedure well.
 
Additional Manufacturer Narrative
H3, h6: it was reported that a left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment have been requested for this complaint but have not become available.A review of the historical complaints data for the acetabular cup and the hemi head and sleeve was performed.Using batch numbers to evaluate patterns of repeated failures or defects over the lifetime of the product and using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware date.No other similar complaints were identified to involve this batch for the acetabular cup and hemi head.Other similar complaints were identified to involve this batch for the sleeve.However, as the device is no longer sold, no action is to be taken.No other similar complaint has been identified for the part number and the reported failure mode for the sleeve.Other similar complaints have been identified for the part number and the reported failure mode for the acetabular cup and hemi head.However, hemi heads, sleeves are no longer sold, no action is to be taken.This failure will continue to be monitored for the acetabular cup.In the absence of the actual devices, the production records were reviewed for the known devices reportedly involved in this incident.All the released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, no prior applicable escalation actions were identified for the acetabular cup.Prior applicable escalation actions were identified for the hemi head and sleeve and confirmed to reduce associated risks as far as possible; no further escalation actions required.The available medical documents were reviewed.Heterotopic ossification is a known complication of joint surgeries and is related to the procedure.The revision operative report did not note any findings of metallosis.The ¿grossly loose¿ acetabular component and cavitary defects of the acetabulum noted intraoperatively are consistent with findings associated with osteolysis.However, based on the information provided, the clinical root cause cannot be confirmed.Although it was reported the patient experienced chromium toxicity, neither the levels nor laboratory reports were provided for review.Without the supporting laboratory/pathology results, imaging, and/or the analysis of the explanted components, the root cause of the reported events and intraoperative findings cannot be confirmed.It cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant or implant failure.The patient impact beyond the revision cannot be determined.Based on the information provided, we can confirm this complaint.Heterotopic ossification is a known complication of joint surgeries and a possible root cause in addition to the noted osteolysis and loose acetabular component.However, without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, excessive patient weight, trauma to the joint replacement.Should the devices or additional information be received, the complaint will be reopened.Based on the investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETABULAR CUP HAP SIZE 48/54
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14436754
Report Number3005975929-2022-00277
Device Sequence Number5202925
Product Code NXT
UDI-Device Identifier03596010552297
UDI-Public03596010552297
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/28/2014
Device Model Number74122154
Device Catalogue Number74122154
Device Lot Number09BW21873
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer08/04/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEMI HEAD 48MM; MODULAR SLEEVE -4MM 12/14; SYN POR FEM COMP SZ 15
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-