• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD LEFT SIZE 7; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD LEFT SIZE 7; PROSTHESIS, KNEE Back to Search Results
Model Number 42-5026-062-01
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Noise, Audible (3273); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Swelling/ Edema (4577)
Date of Event 12/24/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent an initial left total knee arthroplasty without complications.Subsequently, the patient has reported ongoing pain, swelling, and popping in the knee for the past two years and four months since implantation.Additionally, the patient is having range of motion issues with one knee and cannot bend or kneel.The patient has had bilateral knee replacements and it is unknown which knee is having range of motion complications.No medical intervention has been reported at the time of this report.It was reported that no further information is available.
 
Additional Manufacturer Narrative
(b)(4).Medical products: natural tibia cemented 5 degree stemmed left size e: catalog#42532007101, lot#64417330; articular surface fixed bearing ultracongruent (uc) left 14 mm height: catalog#42512200514, lot#64279930; all-poly patella cemented 32 mm diameter: catalog#42540200032, lot#64540883; refobacin bc r 1x40 us: catalog#110034355, lot#818eak2709, qty#2.Multiple mdr reports have been filed for this event.Please see associated reports: 3007963827-2022-00137; 3007963827-2022-00138; 0002648920-2022-00124.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b2; b4; b5; g3; h2; h3; h6; h10.H6: proposed component (annex g) code is mechanical (g04) - femur.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.Device history record was reviewed and no discrepancies related to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified no complication during the initial surgery.No further records were provided for ongoing reported issue.Root cause was unable to be determined.Per package insert, pain, poor range of motion, and swelling are known adverse effects of the system.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information at time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD LEFT SIZE 7
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14437452
Report Number3007963827-2022-00136
Device Sequence Number5579160
Product Code MBH
UDI-Device Identifier00889024230354
UDI-Public(01)00889024230354(17)290131(10)64228083
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number42-5026-062-01
Device Catalogue Number42502606201
Device Lot Number64228083
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer06/15/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-