• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL FAST-CATH HEMOSTASIS INTRODUCER; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL FAST-CATH HEMOSTASIS INTRODUCER; INTRODUCER, CATHETER Back to Search Results
Model Number 406124
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
Following the atrial fibrillation procedure, when the sheath was removed from the patient, it was noted that there was a serration type appearance at the edge.
 
Additional Manufacturer Narrative
One image was submitted for evaluation to product performance engineering; no device components were returned.The photos appeared to show multiple locations in the sheath tubing that appeared to have been torn or punctured.Visual inspection was based solely upon a review of the photographs provided.The device was not returned.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH HEMOSTASIS INTRODUCER
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14437654
Report Number3005334138-2022-00333
Device Sequence Number5151294
Product Code DYB
UDI-Device Identifier05414734014052
UDI-Public05414734014052
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K914090
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model Number406124
Device Catalogue Number406124
Device Lot Number8277584
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-