• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBE; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBE; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367844
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 7.2mg blood collection tube the stopper pulled out of the tube when in the analyzer.The following information was provided by the initial reporter.The customer stated: "there was poor closure of the lid that prevents the proper use of the product in the analyzer equipment.".
 
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 7.2mg blood collection tube the stopper pulled out of the tube when in the analyzer.The following information was provided by the initial reporter.The customer stated: "there was poor closure of the lid that prevents the proper use of the product in the analyzer equipment.".
 
Additional Manufacturer Narrative
H.6.Investigation: bd received 3 photos were received in support of this complaint.The photos do show lavender stopper particles in a filter being held and lavender stopper particles on a machine part, also a photo showing tubes in shelf packs.It cannot be seen if the particles came from actual coring of the stopper and, no photos show an actual stopper after needle puncture.As no samples were received, the investigation is limited.Therefore, 10 production lot in-house retention tubes were inspected with 0 visible defects and.Functional testing was performed at the manufacturing site on the 10-production lot in-house retention tubes.All tubes were within specification limits and no coring was observed.Bd was unable to confirm the customer¿s indicated failure mode no samples were received, and the defect was not observed in the photos or the retention sample testing.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBE
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14437994
Report Number1917413-2022-00276
Device Sequence Number17575954
Product Code JKA
UDI-Device Identifier50382903678445
UDI-Public50382903678445
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
BK050036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2022
Device Model Number367844
Device Catalogue Number367844
Device Lot Number1098448
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-