• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY SO POROUS SIZE 6; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY SO POROUS SIZE 6; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71356006
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Infiltration into Tissue (1931); Synovitis (2094)
Date of Event 10/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, after a left bhr-tha was performed on (b)(6) 2010 due to degenerative arthritis, the patient experienced increasing cobalt and chromium levels.An mri showed fluid collection with synovial thickening, consistent with adverse local tissue reaction.This adverse event was treated with a left revision surgery on (b)(6) 2021.Patient's current health status is unknown.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, with the information provided the clinical root cause of the reported ¿increasing metal ions and moderate synovitis¿ cannot be confirmed.It cannot be concluded the reported clinical reactions were associated with a malperformance of the implant or implant failure.The patient impact beyond the revision and expected transient post-op convalescence period cannot be determined.No further clinical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the instructions for use for this device revealed that metal sensitivity reactions has been identified as an adverse event in primary and revision surgery.Also, the postoperative section establishes that the patient should be advised to report any unusual incidences.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include damaged product, implant corrosion or wear.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY SO POROUS SIZE 6
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14438332
Report Number1020279-2022-02430
Device Sequence Number5634901
Product Code JDI
UDI-Device Identifier03596010567765
UDI-Public03596010567765
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P030022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/25/2020
Device Model Number71356006
Device Catalogue Number71356006
Device Lot Number10DM15415
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date06/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/28/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACETLR CUP HAP 52MM W/ IMPTR.; HEMI HEAD 46MM.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-