|
Related manufacturer reference number: 2017865-2022-10449.It was reported that the patient presented to the emergency room.During device interrogation, it was discovered that the left ventricular(lv) and right ventricular(rv) leads were dislodged.The dislodgment was confirmed via x-ray imaging.The patient presented to a scheduled lead revision.During the procedure, the rv lead could not be re-positioned as the helix would not extend.The rv and lv leads were explanted and replaced.The patient condition was stable post-procedure.
|