SMITH & NEPHEW, INC. AWL TAPERED 3.8MM; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72201915
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a tendon transposition procedure, the tapered awl was disassembled in the surgeon's hand when it was tried to remove it from the patient.Force was used and it still did not come off the bone, requiring the use of pliers and a hammer to remove it.All the pieces were removed from the patient.It is unknown how the procedure was completed.No surgical delay nor further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Per case details, the tapered awl was reportedly, disassembled in the surgeon¿s hand, during an attempt to remove it from the patient.Per case correspondence, after the surgeon unsuccessfully applied force to remove the device, ¿pliers and a hammer¿ was used to remove the device.¿it was confirmed that all pieces were removed from the patient.¿ it is unknown how the surgical procedure was completed; however, there was ¿no surgical delay.¿ smith and nephew has not received the requested patient specific documentation, the device, and/or adequate materials to fully evaluate the root cause of the complaint.Patient impact beyond that which was reported could not be determined.There was no way to determine if the device contributed to the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The handle at the proximal end of the awl has fractured away and was not returned.The awl shows wear from use down the shaft.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with unintended use of the device.Factors that could have contributed to the failure include an application of unintended inappropriate or excessive force to the device, an impact event inconsistent with normal use, misuse or rough handling.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AWL TAPERED 3.8MM
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14438545
Report Number1219602-2022-00754
Device Sequence Number17557453
Product Code NBH
UDI-Device Identifier03596010610607
UDI-Public03596010610607
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72201915
Device Catalogue Number72201915
Device Lot Number50378093
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer07/13/2022
08/25/2022
01/05/2024
01/05/2024
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date07/14/2022
08/26/2022
01/08/2024
01/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/24/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age31 YR
Patient SexMale
-
-