Model Number 71453232
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference number: (b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery had been performed, the patient had a fall causing the post to fracture.This adverse event was addressed by a revision surgery on (b)(6) 2022 to exchange the lgn ps high flex xlpe sz 7-8 11mm.Current health status of patient is unknown.
 
Additional Manufacturer Narrative
H10: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, an insert revision was performed after an unknown length of time in-situ¿ due to the surgeon ¿suspected that the post was fractured¿ after the ¿patient had a fall¿.The requested medical documentation has not been received as of the date of this medical investigation.The current health status of patient remains unknown.Without the requested clinical documentation, the root cause of the reported event could not be further assessed.The patient impact beyond the reported fall and subsequent insert revision could not be determined.No further medical assessment could be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that fracture of implant components has been identified as a possible adverse effect that may occur as a result of trauma or strenuous activity.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or potential causes that could contribute to the reported event include excessive forces applied to implant and traumatic injury.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be fractured along the post, rendering the device inoperative.The device's fractured post was returned was returned as well.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.The clinical/medical evaluation concluded that per complaint details, an insert revision was performed after an unknown length of time in-situ¿ due to the surgeon ¿suspected that the post was fractured¿ after the ¿patient had a fall¿.The requested medical documentation has not been received as of the date of this medical investigation.The current health status of patient remains unknown.Without the requested clinical documentation, the root cause of the reported event could not be further assessed.The patient impact beyond the reported fall and subsequent insert revision could not be determined.No further medical assessment could be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use document revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.The contribution of the device to the reported incident could be corroborated as the device was broken and required a revision surgery to treat this event.Factors that could contribute to the reported event include traumatic injury, size selected, surgical technique or postoperative care.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 7-8 11MM
Common Device Name
PROSTH, KNEE, PATELLOFE, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14438557
Report Number1020279-2022-02433
Device Sequence Number5733033
Product Code JWH
UDI-Device Identifier00885556035153
UDI-Public00885556035153
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/10/2018
Device Model Number71453232
Device Catalogue Number71453232
Device Lot Number08HM07924
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer06/17/2022
10/06/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date06/20/2022
10/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/12/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;