• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number RLT231214
Medical Device Problem Code Difficult or Delayed Positioning (1157)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 04/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).According to the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and/or require additional intraoperative procedure time or intervention include, but are not limited to:endoprosthesis improper component placement, occlusion of native vessel.According to the gore® excluder® aaa endoprosthesis instructions for use (ifu) pre-treatment planning: for angiography, use correct imaging angulation (cranial-caudal, lateral-oblique) to accurately identify origin of branch vessel anatomy.
 
Event or Problem Description
On (b)(6) 2022 the patient was admitted for evar (aaa).As reported the device was inserted in the standard fashion, marked the renal arteries, deployed the graft, cannulated, brought the graft down into position under the renal arteries.It was noted that the renal arteries may have been partially covered.The renal arteries seemed to be about equal in position but on ct the right renal was lower.The surgeon elected to cannulate the renal arteries and placed non-gore stents (visi-pro) in order to ensure flow and to prevent any migration of the rlt231214 graft proximally following the procedure.Following this the renal arteries were confirmed to be patent, there were no observed leaks and all stents appeared to be in good positions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
breanna cox
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key14438733
Report Number3007284313-2022-01917
Device Sequence Number12019227
Product Code MIH
UDI-Device Identifier00733132622337
UDI-Public00733132622337
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/26/2024
Device Model NumberRLT231214
Device Catalogue NumberRLT231214
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/25/2022
Initial Report FDA Received Date05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexMale
Patient Weight84 KG
-
-