MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problem
Electromagnetic Interference (1194)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/07/2022 |
Event Type
malfunction
|
Event Description
|
Information was received from a foreign healthcare provider (hcp) via a manufacturer's representative regarding a patient who was receiving 7.5 mg/ml of morphine at 1.8931 mg/day, 25.0 mcg/ml of bupivacaine at 6.310 mcg/day, and 75.0 mcg/ml of clonidine at 18.931 mcg/day via an implantable pump.On (b)(6) 2022, it was reported that a service code 99 (indicative of tube set) and a service code 100 (indicative of a motor stall) was observed on the physician's programmer during a refill.The pump logs were read and the motor stall was noted on (b)(6) 2022 at 15:37, with the pump reaching tube set on (b)(6) 2022 at 15:37.The pump recovered on (b)(6) 2022 when the pump was read at 11:55.It was noted that the patient possibly had an mri on (b)(6) 2022.This information had yet to be confirmed.No further action was taken as the issue was considered resolved at the time of the report.The patient's status was listed as "alive - no injury".
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received.An mri did occur on 2022-apr-07.
|
|