Catalog Number 326668 |
Device Problem
Volume Accuracy Problem (1675)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd ultra-fine¿ insulin syringe scale marking faded.The following information was provided by the initial reporter translated from (b)(6) to english: "according to the customer's report, there was a syringe on which the printed scale was faded.".
|
|
Manufacturer Narrative
|
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 25-may-2022.H.6.Investigation: customer returned (1) 1/2cc, 8mm, 30g syringe in an open poly bag from lot # 1109961.Customer states that there was a syringe on which the printed scale was faded.The returned syringe was examined and no scale markings were printed on the barrel.A review of the device history record was completed for batch# 1109961.All inspections and challenges were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.A definitive root cause cannot be determined.
|
|
Event Description
|
It was reported that the bd ultra-fine¿ insulin syringe scale marking faded.The following information was provided by the initial reporter translated from japanese to english: "according to the customer's report, there was a syringe on which the printed scale was faded.".
|
|