• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY EMMA (MMHG) CAPNOGRAPH KIT; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY EMMA (MMHG) CAPNOGRAPH KIT; OXIMETER Back to Search Results
Model Number 3639
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the investigation facility to allow for an analysis to be performed.If the product is received for evaluation or new information is obtained, a follow up report will be submitted.(b)(6).
 
Event or Problem Description
The customer reported no sound from the speaker.No consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
Additional manufacturing narrative: the returned device was evaluated.During functionality testing, all alarm conditions were visual but not audible due to the speaker.A known good speaker and emitter board was installed into the device and the speaker was audible.A service history record review reveals that this unit was in the field for over four (4) years with no previous reported issues related to this reported event.Initial reporter phone # is entered as (b)(6) as it exceeded the maximum allowable characters.Initial reported phone # is: (b)(6).
 
Event or Problem Description
The customer reported no sound from the speaker.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMMA (MMHG) CAPNOGRAPH KIT
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key14440176
Report Number3019388613-2022-00115
Device Sequence Number17554213
Product Code CCK
UDI-Device Identifier07350046131750
UDI-Public07350046131750
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K063167
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3639
Device Catalogue Number3639
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date06/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/13/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-