| Model Number |
3639 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the investigation facility to allow for an analysis to be performed.If the product is received for evaluation or new information is obtained, a follow up report will be submitted.(b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported no sound from the speaker.No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned device was evaluated.During functionality testing, all alarm conditions were visual but not audible due to the speaker.A known good speaker and emitter board was installed into the device and the speaker was audible.A service history record review reveals that this unit was in the field for over four (4) years with no previous reported issues related to this reported event.Initial reporter phone # is entered as (b)(6) as it exceeded the maximum allowable characters.Initial reported phone # is: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported no sound from the speaker.No consequences or impact to patient were reported.
|
| |
|