Model Number PM2134
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented to the hospital for a pacemaker implant procedure on (b)(6) 2022.While attempting to implant the pacemaker, it was noted that there was set screw issue.The set screw could not be tightened and has a loose connection.After multiple failed attempts, the whole system was removed without further incident in the same procedure.The patient was in stable condition.
 
Additional Manufacturer Narrative
The reported event of the wrench did not click when tightening the ventricular setscrew was confirmed.Analysis revealed the user of the torque wrench was incorrectly inserting the wrench at angles instead of straight down and also not aligning the wrench to go into the hex setscrew inset.The multiple incorrect wrench insertion attempts caused excessive septum punch-outs that landed in and filled the setscrew inset.The excess septum punch-outs limited wrench insertion into the setscrew inset and caused the wrench to slip and strip portions of the inset.Under these conditions caused by the user, the setscrew could not be tightened which means the wrench will not click.The problem is related to the user's incorrect use of the torque wrench.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUSTAIN XL DC
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14442505
Report Number2017865-2022-10469
Device Sequence Number4310479
Product Code NVZ
UDI-Device Identifier05414734504720
UDI-Public05414734504720
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model NumberPM2134
Device Catalogue NumberPM2134
Device Lot NumberP000124430
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2022
Supplement Date Received by Manufacturer08/15/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TENDRIL; TENDRIL
Patient SexUnknown
-
-