COOK INC GUNTHER TULIP JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
| Model Number |
G33017 |
| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Codes |
Internal Organ Perforation (1987); Perforation of Vessels (2135); Anxiety (2328)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Customer (person) not provided, information provided by (b)(6) llp.Occupation: non-healthcare professional investigation: filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Rpn/lot is unknown, however, the device is manufactured and inspected according to current controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report includes information known at this time.A follow up report will be submitted should additional information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
It is alleged that "patient received a cook tulip filter.Patient is alleging filter perforation of organ.Hospital and medical records have not been provided.".
|
| |
|
Event or Problem Description
|
|
Patient allegedly received an implant on (b)(6) 2012 via the right internal jugular vein due to lower extremity deep vein thrombosis (dvt).Patient is alleging vena cava perforation with organ abutment.Patient further alleges anxiety, worry, fear, and limited physical activity.Report from ct (computed tomography): "lines, tubes and hardware: there is an inferior vena cava filter in place with filter apex located just cranial to the right renal vein without significant tilt of the filter.The distal aspect of all 4 feet of the filter appear to extend outside the lumen of the ivc with the anteromedial strut directly abutting the aorta without evidence of penetration as can be seen on this noncontrast exam.The features do not abut the adjacent spinal periosteal or duodenum.".
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: investigation investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: organ/vena cava (vc) perforation, anxiety, worry, fear, limited physical activity.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.Unknown if the reported anxiety, worry, fear, and limited physical activity are directly related to the filter and unable to identify a corresponding failure mode at this point in time.(b)(4) in lot.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|