• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC GUNTHER TULIP JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC GUNTHER TULIP JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G33017
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code Perforation of Vessels (2135)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(6).Occupation: non-healthcare professional.Investigation: filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Rpn/lot is unknown, however, the device is manufactured and inspected according to current controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report includes information known at this time.A follow up report will be submitted should additional information become available.
 
Event or Problem Description
It is alleged that "patient received a cook tulip filter.Patient is alleging perforation of ivc.Hospital and medical records have not been provided.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP JUGULAR VENA CAVA FILTER SET
Common Device Name
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14444184
Report Number1820334-2022-00882
Device Sequence Number6354285
Product Code DTK
UDI-Device Identifier10827002330174
UDI-Public(01)10827002330174(17)150601(10)3529445
Combination Product (Y/N)N
PMA/510(K) Number
K072240
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG33017
Device Catalogue NumberIGTCFS-65-JUG
Device Lot Number3529445
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/20/2022
Initial Report FDA Received Date05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/26/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexFemale
-
-