Customer (person) not provided, information provided by onderlaw, llc.Occupation: non-healthcare professional.Investigation: filter fracture has been reported and may be either symptomatic or asymptomatic.Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e.G., a renal vein).Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported.Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures.This report includes information known at this time.A follow up report will be submitted should additional information become available.
|