Model Number G52909
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Codes Internal Organ Perforation (1987); Perforation of Vessels (2135); Anxiety (2328)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer (person) not provided, information provided by (b)(6).Occupation: non-healthcare professional.Investigation: filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Rpn/lot is unknown, however, the device is manufactured and inspected according to current controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report includes information known at this time.A follow up report will be submitted should additional information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It is alleged that "patient received a cook celect filter.Patient is alleging perforation of organ.Hospital and medical records have not been provided.".
 
Event or Problem Description
The patient alleges inferior vena cava perforation, one prong abuts the osteophyte at l3-4.
 
Additional Manufacturer Narrative
Investigation: the following allegations have been investigated: organ/vena cava perforation.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.20 devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: anxiety, worry, fear, limited physical activity.Unknown if the reported anxiety, worry, fear, and limited physical activity are directly related to the filter and unable to identify a corresponding failure mode at this point in time.(b)(4) in lot.Three other complaints have been reported against the lot; one for a different issue and two for similar issues.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Patient allegedly received a filter via the right internal jugular vein due to deep vein thrombosis (dvt).Patient is alleging vena cava perforation abutting another organ.Patient notes and further alleges experiencing anxiety, worry, fear, limited physical activity.Computed tomography (ct) abdomen/pelvis: "cook celect filter placed 11/13/2012"."caval perforation: yes.Grade 3 perforation of 6 o'clock strut to abut osteophyte at l3-4.Grade 2 perforation of 9 o'clock strut 0.58 cm, 12 o'clock strut 0.46 cm and 3 o'clock strut 0.63 cm.Tilt: no substantial tilt.Apex of filter does not touch wall of ivc.Migration: no.Pertinent negatives: no definite fracture or missing struts.No ivc stenosis.Additional findings: no".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVALIGN COOK CELECT JUGULAR VENA CAVA FILTER SET
Common Device Name
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
henriette s. christiansen
sandet 6, dk-4632
bjaeverskov 
MDR Report Key14444756
Report Number3002808486-2022-00652
Device Sequence Number14530811
Product Code DTK
UDI-Device Identifier10827002529097
UDI-Public(01)10827002529097(17)150910(10)E2975802
Combination Product (Y/N)N
PMA/510(K) Number
K090140
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG52909
Device Catalogue NumberIGTCFS-65-1-JUG-CELECT
Device Lot NumberE2975802
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/20/2022
Supplement Date Received by Manufacturer03/15/2023
01/09/2024
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date03/23/2023
02/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/11/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexMale
Patient Weight88 KG
-
-