|
WILLIAM COOK EUROPE COOK CELECT FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
| Model Number |
G48373 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Structural Problem (2506); Device Tipped Over (2589); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Internal Organ Perforation (1987); Pain (1994); Perforation of Vessels (2135); Thromboembolism (2654); Device Embedded In Tissue or Plaque (3165); Thrombosis/Thrombus (4440); Unspecified Gastrointestinal Problem (4491); Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It is alleged that "patient received a cook celect filter.Patient is alleging fracture.Hospital and medical records have not been provided.".
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter address: customer (person) not provided, information provided by onderlaw, llc.Initial reporter occupation: non-healthcare professional.Investigation: filter fracture has been reported and may be either symptomatic or asymptomatic.Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e.G., a renal vein).Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported.Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures.This report includes information known at this time.A follow up report will be submitted should additional information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information: d4 (expiration date), h6 (device problem line #1) corrected from a0511-structural problem to a0501-detachment of device or device component investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: fracture, organ/vena cava (vc) perforation, embedment, tilt.Filter fracture has been reported and may be either symptomatic or asymptomatic.Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e.G., a renal vein).Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported.Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.(b)(4) devices in lot.One other complaint, for a different issue, has been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Patient allegedly received a cook celect inferior vena cava (ivc) filter via the right common femoral vein.Per a report from computed tomography (ct): " caval perforation: yes.12 o'clock 11.50mm grade 4 extending into the duodenum.9 o'clock 11.20mm grade 2.6 o'clock 20.80mm grade 4 fractured strut extending into the l3 vertebral body.3 o'clock 3.20mm grade 3 extending to the periphery of the abdominal aorta.Tilt: yes.Left sided tilt measuring 15.64 degrees.Migration: no.Pertinent negative: no free air, no hematoma.Additional findings: fractured posterior strut fragment embedded into the l3 vertebral body as described".
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information: d4 (model #).Additional information: a4, b5, b6, b7, h6.Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: deep vein thrombosis (dvt), pain, back pain, gastroenteritis, inability to work, limited physical activity.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Unknown if the reported pain, back pain, gastroenteritis, inability to work, and limited physical activity are directly related to the filter and unable to identify a corresponding failure mode at this point in time.20 devices in lot.One other complaint has been reported against the lot for different issue.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Patient allegedly received the cook celect filter due to deep vein thrombosis (dvt).Patient is alleging device tilt - left sided tilt measuring 15.64 degrees, vena cava perforation - grade 2, 3 perforations; multiple grade 4 perforations, fracture - fragment embedded in the l3 vertebral body, organ perforation into the duodenum and l3 vertebral body.Patient notes and further alleges experiencing several dvt post filter placement, pain, back pain, gastroenteritis issues, inability to work or take walks with family; limited physical activity due to pain related to the filter.
|
| |
|
Search Alerts/Recalls
|
|
|