Catalog Number 8065797105
Medical Device Problem Codes Gas Output Problem (1266); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A customer reported that perfluoropropane gas bubble in eye reduced in size prematurely in the patient eye.Procedure was completed and patient harm was not reported.Additional information has been received indicating that the event occurred during surgery but not noted until post operative (post-op).The gas fill was noted to be low at post-op and perfluoropropane was not dispensed.
 
Additional Manufacturer Narrative
The c3f8 (perfluoropropane) tank was not returned for evaluation.A review of the batch production record could not be performed because the lot number was not reported.The expiration date could not be determined.A review of the complaint records could not be performed because the lot number was not provided.A review of confirmed complaints for leaking cylinders of this type or for c3f8 not lasting as long as expected.Both were for cylinders that leaked at the valve to cylinder connection.The root cause of the reported event cannot be determined.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate the manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISPAN PERFLUOROPROPANE (C3F8) GAS
Common Device Name
INTRAOCULAR GAS
Manufacturer (Section D)
AIRGAS THERAPEUTICS LLC
6141 easton road
p.o. box 310
plumsteadville PA 18949
Manufacturer (Section G)
AIRGAS THERAPEUTICS LLC
6141 easton road
p.o. box 310
plumsteadville PA 18949
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14445076
Report Number1610287-2022-00027
Device Sequence Number19963203
Product Code LPO
UDI-Device Identifier00380657971053
UDI-Public00380657971053
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P900066
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065797105
Device Lot NumberASKU
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/20/2022
Supplement Date Received by Manufacturer09/02/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONSTELLATION VISION SYSTEM
Patient SexMale
-
-